[en] Background
The gold standard for post-operative deep brain stimulation (DBS) parameter tuning is a monopolar review of all stimulation contacts, a strategy being challenged by recent developments of more complex electrode leads.
Objective
Providing a method to guide clinicians on DBS assessment and parameter tuning by automatically integrating patient individual data.
Methods
We present a fully automatic method for visualization of individual deep brain structures in relation to a DBS lead by combining precise electrode recovery from post-operative imaging with individual estimates of deep brain morphology utilizing a 7T-MRI deep brain atlas.
Results
The method was evaluated on 20 STN DBS cases. It demonstrated robust automatic creation of 3D-enabled PDF reports visualizing electrode to brain structure relations and proved valuable in detecting miss placed electrodes.
Discussion
Automatic DBS assessment is feasible and can conveniently provide clinicians with relevant information on DBS contact positions in relation to important anatomical structures.
Research center :
- Luxembourg Centre for Systems Biomedicine (LCSB): Systems Control (Goncalves Group)
Disciplines :
Human health sciences: Multidisciplinary, general & others
Author, co-author :
Husch, Andreas ; University of Luxembourg > Luxembourg Centre for Systems Biomedicine (LCSB)
Petersen, Mikkel V.; Aarhus Universitet - AU > Department of Clinical Medicine - Center of Functionally Integrative Neuroscience,
Gemmar, Peter
Goncalves, Jorge ; University of Luxembourg > Luxembourg Centre for Systems Biomedicine (LCSB)
Sunde, Niels
Hertel, Frank ; University of Luxembourg > Luxembourg Centre for Systems Biomedicine (LCSB)
External co-authors :
yes
Language :
English
Title :
Post-operative deep brain stimulation assessment: Automatic data integration and report generation