Can the length of a catheter change the time to bubble at the tip performing the “Bubble Test”? A bench study
Giustivi, Davide; Elli, Stefano; Airoldi, Chiara et al.
2025In Journal of Vascular Access, 26 (1), p. 109 - 115
Peer Reviewed verified by ORBi
 

Files


Full Text
giustivi-et-al-2023-can-the-length-of-a-catheter-change-the-time-to-bubble-at-the-tip-performing-the-bubble-test-a.pdf
Author postprint (549.93 kB)
Request a copy

All documents in ORBilu are protected by a user license.

Send to



Details



Keywords :
Bubble test; central venous catheters; PICC; RASS; tip location; Catheterization, Central Venous; Catheters, Indwelling; Central Venous Catheters; Clinical Competence; Contrast Media; Equipment Design; Humans; Materials Testing; Predictive Value of Tests; Time Factors; Ultrasonography, Interventional; contrast medium; Article; bubble test; contrast-enhanced ultrasound; electrocardiography; flow rate; laboratory test; central venous catheter; central venous catheterization; clinical competence; comparative study; devices; equipment design; human; indwelling catheter; interventional ultrasonography; materials testing; predictive value; time factor
Abstract :
[en] Introduction: Intraprocedural tip control techniques are critical during central venous catheter placement. According to international guidelines (INS 2021), intracavitary electrocardiography is the first method of choice to verify it; when this technique is not feasible, it is considered acceptable to use a contrast-enhanced ultrasound-based tip location method, commonly known as “bubble-test” as an effective alternative. Objective: To assess whether the length of the vascular catheter can alter the time between the injection of the contrast media and its appearance at the catheter tip and the injection duration. Differences between operators stratified according to experience were evaluated as secondary endpoints. Methods: A bench study was conducted using an extracorporeal circuit. For each catheter length (60, 40, and 20 cm), three injections were obtained by each of the five operators with different levels of experience for a total of 45 measurements. Differences among operators were evaluated using ANOVA, and the impact of catheter length and operator expertise on times was assessed using repeated measurement models. Results: Hub-to-tip times of 247.33 ms (SD 168.82), 166 ms (SD 95.46), 138 ms (SD 54.48), and injection duration of 1620 ms (SD 748.58), 1614 ms (SD 570.95), 1566 ms (SD 302.83) were observed for 60, 40, 20 cm catheter length, respectively. A significant time variability between operators was observed. Moreover, moving from 60 to 20 cm, hub-to-tip time was significantly longer for 60 cm devices (p = 0.0124), while little differences were observed for injection duration. Conclusions: Catheter length can change both the time between the injection of the contrast media and its appearance at the catheter tip and the injection duration. Hub-to-tip times obtained with 20 and 40 cm and overall injection duration did not differ significantly; skilled personnel could substantially reduce both values analyzed in this study. © The Author(s) 2023.
Disciplines :
Nursing Science
Author, co-author :
Giustivi, Davide;  Vascular Access Team, ASST Lodi, Italy
Elli, Stefano;  Vascular Access Team, IRCCS S.Gerardo dei Tintori, Monza, Italy
Airoldi, Chiara;  Department of Translational Medicine, Università del Piemonte Orientale, Novara, Italy
Lo Izzo, Federica;  Paediatric Departement, ASST Lodi, Italy
Rossini, Michela;  Department of Emergency Medicine, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy
Gidaro, Antonio;  Department of Biomedical and Clinical Sciences “Luigi Sacco,”, University of Milan, Luigi Sacco Hospital, Milan, Italy
Lucchini, Alberto;  General Intensive Care Unit, IRCCS S.Gerardo dei Tintori, Monza, Italy ; Bicocca University, Milan, Italy
PRIVITERA, Daniele  ;  University of Luxembourg > Faculty of Science, Technology and Medicine (FSTM) > Department of Life Sciences and Medicine (DLSM) > Medical Education ; Department of Emergency Medicine, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy
External co-authors :
no
Language :
English
Title :
Can the length of a catheter change the time to bubble at the tip performing the “Bubble Test”? A bench study
Publication date :
2025
Journal title :
Journal of Vascular Access
ISSN :
1129-7298
eISSN :
1724-6032
Publisher :
SAGE Publications Ltd
Volume :
26
Issue :
1
Pages :
109 - 115
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBilu :
since 26 November 2025