Numerical comparative study of five currently used implants for high tibial osteotomy: realistic loading including muscle forces versus simplified experimental loading
DIFFO KAZE, Arnaud; MAAS, Stefan; KEDZIORA, Slawomir et al.
2017In Journal of Experimental Orthopaedics, 5 (28)
Peer Reviewed verified by ORBi
 

Files


Full Text
DiffoKaze_et_al_2017.pdf
Publisher postprint (4.37 MB)
Request a copy

All documents in ORBilu are protected by a user license.

Send to



Details



Keywords :
Finite element; Musculoskeletal model; Lower limb; Stance phase; Medial open wedge HTO; iBalance; Contour lock; PEEKPower; TomoFix; Section forces
Abstract :
[en] Background Many different fixation devices are used to maintain the correction angle after medial open wedge high tibial osteotomy (MOWHTO). Each device must provide at least sufficient mechanical stability to avoid loss of correction and unwanted fracture of the contralateral cortex until the bone heals. In the present study, the mechanical stability of following different implants was compared: the TomoFix small stature (sm), the TomoFix standard (std), the Contour Lock, the iBalance and the second generation PEEKPower. Simplified loading, usually consisting of a vertical load applied to the tibia plateau, is used for experimental testing of fixation devices and also in numerical studies. Therefore, this study additionally compared this simplified experimental loading with a more realistic loading that includes the muscle forces. Method Two types of finite element models, according to the considered loading, were created. The first type numerically simulated the static tests of MOWHTO implants performed in a previous experimental biomechanical study, by applying a vertical compressive load perpendicularly to the plateau of the osteotomized tibia. The second type included muscle forces in finite element models of the lower limb with osteotomized tibiae and simulated the stance phase of normal gait. Section forces in the models were determined and compared. Stresses in the implants and contralateral cortex, and micromovements of the osteotomy wedge, were calculated. Results For both loading types, the stresses in the implants were lower than the threshold values defined by the material strength. The stresses in the lateral cortex were smaller than the ultimate tensile strength of the cortical bone. The implants iBalance and Contour Lock allowed the smallest micromovements of the wedge, while the PEEKPower allowed the highest. There was a correlation between the micromovements of the wedge, obtained for the simplified loading of the tibia, and the more realistic loading of the lower limb at 15% of the gait cycle (Pearson’s value r = 0.982). Conclusions An axial compressive load applied perpendicularly to the tibia plateau, with a magnitude equal to the first peak value of the knee joint contact forces, corresponds quite well to a realistic loading of the tibia during the stance phase of normal gait (at 15% of the gait cycle and a knee flexion of about 22 degrees). However, this magnitude of the knee joint contact forces overloads the tibia compared to more realistic calculations, where the muscle forces are considered. The iBalance and Contour Lock implants provide higher rigidity to the bone-implant constructs compared to the TomoFix and the PEEKPower plates.
Disciplines :
Mechanical engineering
Author, co-author :
DIFFO KAZE, Arnaud ;  University of Luxembourg > Faculty of Science, Technology and Communication (FSTC) > Engineering Research Unit
MAAS, Stefan ;  University of Luxembourg > Faculty of Science, Technology and Communication (FSTC) > Engineering Research Unit
KEDZIORA, Slawomir  ;  University of Luxembourg > Faculty of Science, Technology and Communication (FSTC) > Engineering Research Unit
Belsey, James
Haupert
WOLF, Claude ;  University of Luxembourg > Faculty of Science, Technology and Communication (FSTC) > Engineering Research Unit
Hoffmann, Alexander
Pape, Dietrich
External co-authors :
yes
Language :
English
Title :
Numerical comparative study of five currently used implants for high tibial osteotomy: realistic loading including muscle forces versus simplified experimental loading
Publication date :
06 November 2017
Journal title :
Journal of Experimental Orthopaedics
eISSN :
2197-1153
Publisher :
Springer International Publishing A.G., Heidelberg, Switzerland
Volume :
5
Issue :
28
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBilu :
since 29 March 2019

Statistics


Number of views
117 (19 by Unilu)
Number of downloads
0 (0 by Unilu)

Scopus citations®
 
7
Scopus citations®
without self-citations
6
OpenCitations
 
2

Bibliography


Similar publications



Contact ORBilu