See detailUse of the Sideguard (Cappella) stent in bifurcation lesions: a real-world experience.
Mamas, Mamas A.; Farooq, Vasim; Latib, Azeem et al

in EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology (2012), 7(10), 1170-80

AIMS: The Sideguard(R) stent (Cappella Medical Devices Ltd, Galway, Ireland), is a novel nitinol self-expanding dedicated bifurcation stent that flares proximally at the ostium of the side branch (SB ... [more ▼]

AIMS: The Sideguard(R) stent (Cappella Medical Devices Ltd, Galway, Ireland), is a novel nitinol self-expanding dedicated bifurcation stent that flares proximally at the ostium of the side branch (SB) into a trumpet shape thereby achieving full ostial coverage. The aim of this study is to report the utility and limitations of this stent in patients undergoing treatment to bifurcation coronary lesions in a real-world setting. METHODS AND RESULTS: We prospectively identified 20 successive patients admitted over a 6-month period in whom there was significant SB disease and who were suitable for a bifurcation procedure. The Sideguard(R) stent was successfully used in all 20 cases including several that would have been technically difficult using conventional bifurcation techniques. We highlight use of this system using five illustrative cases that illustrate its utility and limitations in the treatment of bifurcation lesions. CONCLUSIONS: The Sideguard(R) stent can be used to treat complex bifurcation lesions in a straight forward manner and is not subject to the limitations associated with conventional bifurcation PCI techniques including jailing of the SB ostium and inability to fully cover/scaffold the ostium of the SB. [less ▲]

Detailed reference viewed: 111 (0 UL)
Full Text
Peer Reviewed
See detailUse of the sheathless guide catheter during routine transradial percutaneous coronary intervention: a feasibility study.
Mamas, Mamas; D'Souza, Savio; Hendry, Cara et al

in Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2010), 75(4), 596-602

OBJECTIVE: The aim of this study is to investigate the feasibility of using a 6.5 Fr sheathless guide catheter as a default system in transradial (TRA) percutaneous coronary intervention (PCI). BACKGROUND ... [more ▼]

OBJECTIVE: The aim of this study is to investigate the feasibility of using a 6.5 Fr sheathless guide catheter as a default system in transradial (TRA) percutaneous coronary intervention (PCI). BACKGROUND: TRA PCI has been shown to reduce mortality rates through a reduction in access site related bleeding complications compared with procedures performed though a femoral approach. Complications associated with the TRA route increase with the size of sheath used. These complications may be reduced by the use of a sheathless guide catheter system (Asahi Intecc, Japan) that is 1-2 Fr sizes smaller in diameter than the corresponding introducer sheath. METHODS: We performed PCI in 100 consecutive cases using 6.5 Fr sheathless guides to determine the procedural success, rates of symptomatic radial spasm and radial occlusion. RESULTS: Procedural success using the 6.5 Fr sheathless guide catheter system was 100% with no cases requiring conversion to a conventional guide and catheter system. There were no procedural complications recorded associated with the use of the catheter. Adjunctive devices used in this cohort included IVUS, stent delivery catheters, distal protection devices, and simple thrombectomy catheters. The rate of radial spasm was 5% and the rate of radial occlusion at 2 months was 2%. CONCLUSION: Use of the 6.5 Fr sheathless guide catheter system, which has an outer diameter <5 Fr sheath, as the default system in routine PCI is feasible with a high rate of procedural success via the radial artery. [less ▲]

Detailed reference viewed: 124 (0 UL)